Target: Flow
Model: FR15
Detection principle: MEMS thermal
Characteristics: High sensitivity, very low start-up flows; High response time
Response time: ≤3ms

FR15 Manual download
Get Price Whatsapp

Description

The FR15 gas flow sensor is specially developed for the medical field. It uses the MEMS thermal principle to flow the vascular gas medium and is suitable for critical ventilators/transport ventilators, etc. The product integrates two types of MEMS flow and temperature sensitive components, and is equipped with gas medium mass flow and medium temperature measurement functions. 

Application

see manual

Features

  • High sensitivity, very low start-up flows;
  • High response time;
  • High accuracy and measurement repeatability;
  • Wide range;
  • Biocompatible gas connection material

Specifications

Model No. FR15
D ∅15mm
Flow measurement Maximum flow rate 240L/min @20℃ 101.325kPa
Measurement accuracy 9L/min~240L/min   ±2.5% 0~9L/min        ±0.5%FS
Repeatability 0.5%
Working pressure 70kPa~107kPa
Burst pressure ≥0.7MPa
Working temperature 5℃~55℃
Temperature measurement Measurement range ﹣20℃~+80℃
Measurement accuracy 0℃~80℃   ±2.5℃
﹣20℃~0℃   ±4℃
Output signal Output method Digital IIC and linear analogue voltage
IIC communication rate 400kHz
Digital Signal Refresh Time ≤1ms
Analogue flow linearity 0.25V~2V
Analogue temperature linearity 0.25V~2V
Signal Response TimeT10~T90 ≤3ms
Electrical parameters Working voltage 5 V~14V
Rated power ﹤0.3W
Electrical interface Sensor:Molex 87831-1041 Lead wire:Molex 87832-1010
Others Storage temperature -20℃~80℃
Pressure losses <1800Pa  @200L/min 101.325kPa
Measuring medium Dry, clean, non-corrosive gases
Weight 22g±1g

* The default flow sensor of our company adopts 20℃ 101.325kPa, air calibration, and the production conditions are temperature 22±2℃, purified, (30%~35%) RH environment. If the user has special requirements, it will be calibrated according to the customer's requirements.
*··  %FS refers to full scale accuracy, % is reading accuracy.

* The description and specifications might be changed without notice. Please contact us for the latest information before placing orders.

Technical Support:

Customized technical design and function, LOGO and package are available without extra charge (including molding fee) for mass volume.

The sensors don't have sleeping mode. It can be set in program if needed.

All Winsen products go through strict inspections and tests to make sure they are 100% qualified before shipping out.

The sensor has been fully tested (not calibrated) before leaving the factory, but the module will be calibrated before leaving the factory.

However, after transportation and storage, the environment conditions change, and the surface of the sensor absorbs moisture, mixed gas, pollution matters, etc. It is recommended that customers perform aging and calibration after sensors become stable according to the program and use occasions of the whole device.

15﹪-90﹪RH without condensation,greater or less than this range or long-term high temperature and high humidity will affect the life of the sensor.

Our sensors are certified with CE, 3C, ROHS, anti-explosion, fire prevention, and coal mining safety according to different applications. And we provide assistance for new certification.

Many factors can cause the sensor to fail such as broken wires in transportation, poisoning, falling off of the sensitive material, damage to the heating wire (device), etc. Please find specific cause in each category with Winsen’s aftersales.

Winsen provides professional consultation and service before and after sales. We will recommend the best fitting sensor for your demand. If you have any problem, don’t hesitate to email us to sales@winsensor.com

Due to high volume of daily incoming emails and time zone difference, we may not be able to reply your emails immediately. But we will reply to your inquiry in 24 business hours.

LEADING GAS SENSING SOLUTION SUPPLIER